Model 5014i to 5030i Upgrade Procedures
SHARP External Valve Installation
10-8 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
c. Connect the 3/8” tubing with inline HEPA filter to the available
port (Port #1) on the external valve assembly using the available
push connect fitting.
Note This step can also be delayed until after the zero air delivery tube and
T-fitting are installed at the sample intake.
d. At the sample inlet, attach the stainless steel T-fitting to the sample
downtube and tighten the swagelok fitting with Teflon ferrules—
do not overtighten these ferrules.
e. Snake the green vacuum tubing downwards towards the rear valve
assembly and attach to Port #1 of the valve. If this is a field
installation, proper water sealants should be used to avoid any
leakage into the instrument.
Figure 10–8. Plumbing Diagram