Servicing
Detector Amplifier Assembly Replacement
7-16 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
2. Remove the SHARP optics cover assembly and the measurement case
assembly, as described earlier.
3. Disconnect the cable from the detector amplifier assembly connector
on the measurement interface board and disconnect the SHV
connector from the detector amplifier.
4. Remove the one screw holding the detector amplifier mounting bracket
and slide out the detector amplifier assembly.
Figure 7–7. Replacing the Detector Amplifier Assembly
5. Install the new detector amplifier assembly by following the previous
steps in reverse.
6. Calibrate the detector amplifier board. Refer to the “Detector Amplifier
Board Calibration” procedure that follows.
Slide Out
Unfasten screw and slide
detector amplifier out
Unhook connector from
Interface Board
Unhook SHV connector
Vista de pagina 233