Sharp SYSTEM 8/10 Plus Manual de instrucciones Pagina 150

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 403
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 149
Operation
Service Menu
3-88 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
The Calibrate Barometer Pressure Offset screen allows the user to calibrate
the pressure sensor at offset pressure.
Note The operator should use an independent barometer to measure the
ambient pressure and enter the value on this screen before calibrating.
In the Main Menu, choose Service > Pres/Vacuum Calibration >
Baro Pres Calibration > Offset.
CAL BARO PRES OFFSET:
PRESSURE: 760 mmHg
OFFSET: 747 mmHg?
§¨ MOVE CURSOR
©ª CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
The Calibrate Barometer Pressure Span screen allows the user to view and
pressure sensor calibration span point.
Note The operator should use an independent barometer to measure the
ambient pressure and enter the value on this screen before calibrating.
In the Main Menu, choose Service > Pres/Vacuum Calibration >
Baro Pres Calibration > Span.
CAL BARO PRESSURE SPAN:
PRESSURE: 760 mmHg
SET TO: 747 mmHg?
§¨ MOVE CURSOR
©ª CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
The Restore Default Calibration screen allows the user to reset the default
values.
In the Main Menu, choose Service > Pres/Vacuum Calibration >
Baro Pres Calibration > Set Defaults.
Calibrate Barometer
Pressure Offset
Calibrate Barometer
Pressure Span
Restore Default
Calibration
Vista de pagina 149
1 2 ... 145 146 147 148 149 150 151 152 153 154 155 ... 402 403

Comentarios a estos manuales

Sin comentarios