Servicing
Removing the SHARP Optics Assembly
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-11
The SHARP optics assembly can be removed from the measurement case
assembly. Refer to the following steps when a procedure requires lowering
the partition panel (Figure 7–4).
Figure 7–4. Removing the SHARP Optics Assembly
Equipment Required:
Philips screwdriver
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
1. Turn the instrument OFF and unplug the power cord.
2. If the instrument is mounted in a rack, remove it from the rack.
3. Disconnect the electrical connector.
4. Unfasten the two captive hardware screws.
Removing the SHARP
Optics Assembly
Captive Screws (2)
Electrical Connector