Model 5030i SHARPInstruction Manual Synchronized Hybrid Ambient Real-time Particulate Monitor Part Numbe
About This Manual Where to Get Help vi Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
Operation Volumetric Conditions 3-38 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific INSTRUMENT ID: CURRENTLY: 14
Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-39 Choices are Concentrations, Other Measurements
Operation Volumetric Conditions 3-40 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific CHOOSE STREAM DATA: >CONCENTRATIONS OTHER
Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-41 OTHER MEASUREMENTS: >NONE BARO PRES VAC
Operation Volumetric Conditions 3-42 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ● In the Main Menu, choose Instrument Control
Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-43 The Netmask screen is used to edit the netmas
Operation Volumetric Conditions 3-44 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific HOST NAME: CURRENT: ISERIES IS
Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-45 Note The digital outputs may take up to one s
Operation Volumetric Conditions 3-46 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The Alarms screen allows the user to select a
Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-47 Not all of the I/O available in the instrument
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual vii Contents Introduction ...
Operation Volumetric Conditions 3-48 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific CHOOSE ACTION: >NONE AOU
Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-49 ALLOW OVER/UNDER RANGE: CURRENTLY:
Operation Volumetric Conditions 3-50 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific analog output channel. See Table 3–4 for a lis
Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-51 Output Zero (0)% Value Full-Scale 100% Value
Operation Volumetric Conditions 3-52 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Concentrations Other Measurements Analog Inp
Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-53 ● In the Main Menu, choose Instrument Control
Operation Volumetric Conditions 3-54 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ANALOG INPUT UNITS: CURRENTLY: V
Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-55 The Table Point submenu allows the user to se
Operation Volumetric Conditions 3-56 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific TABLE POINT 01 USER VAL: CURRENTLY:
Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-57 Note The service mode should be turned off wh
Contents viii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Firmware Overview ...
Operation Diagnostics Menu 3-58 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ● In the Main Menu, choose Instrument Controls >
Operation Diagnostics Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-59 PROGRAM VERSIONS: PRODUCT: MODEL 5030i VE
Operation Diagnostics Menu 3-60 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific INTERFACE BOARD VOLTAGES: 3.3 SUPPLY 3.3
Operation Diagnostics Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-61 NEPH BOARD VOLTAGES: 3.3 SUPPLY 3.3 V
Operation Diagnostics Menu 3-62 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific FLOW: 16.67 LPM
Operation Diagnostics Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-63 CRn 0.00 Bq/m3
Operation Diagnostics Menu 3-64 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific DIGITAL INPUTS: >INPUT 1 1 INPU
Operation Diagnostics Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-65 operation upon exiting this screen. The following e
Operation Alarms Menu 3-66 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The alarms menu allows the user to view a set of alarms
Operation Alarms Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-67 INSTRUMENT ALARMS: >FILTER TAPE COUNTER OK FILTE
Contents Thermo Fisher Scientific Model 5030i SHARP Instruction Manual ix Datalogging Settings ...
Operation Alarms Menu 3-68 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific DETECTOR ALARMS: BOARD STATUS OK >ALPHA
Operation Alarms Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-69 temperature, nephelometer RH, LED current, and reference
Operation Alarms Menu 3-70 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific NEPHELOMETER TEMP: ACTUAL MIN: 0.0 oC SET MIN
Operation Alarms Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-71 the minimum or maximum limit, an alarm is activated and
Operation Alarms Menu 3-72 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The RH/Temperature Alarms submenu allows the user to vi
Operation Alarms Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-73 RELATIVE HUMIDITY: ACTUAL MIN: 5.0 % SET
Operation Alarms Menu 3-74 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific alarm (bell) icon appears in the status bar on the Run s
Operation Alarms Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-75 The Minimum Flow Temperature alarm limit screen is used
Operation Alarms Menu 3-76 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific BOARD TEMPERATURE: ACTUAL MIN: 5.0 oC SET MIN
Operation Alarms Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-77 BAROMETRIC PRESSURE: ACTUAL MIN: 400.0 mmHg SET MI
Contents x Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Non-Alarm ...
Operation Alarms Menu 3-78 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The Flow screen displays the current flow reading and s
Operation Alarms Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-79 FLOW ALARMS: >FLOW OK
Operation Alarms Menu 3-80 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific CONC ALARMS: >AVG PM OK AVG SHARP
Operation Alarms Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-81 The Average SHARP screen displays the current average S
Operation Alarms Menu 3-82 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific AVERAGE NEPH CONC: ACTUAL 0.0 >MIN
Operation Alarms Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-83 minimum and maximum instant PM concentration screens fun
Operation Alarms Menu 3-84 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific INSTANT SHARP CONC: ACT MIN: 0.0 SET MIN
Operation Service Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-85 The Service menu appears only when the instrument is i
Operation Service Menu 3-86 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific RH/TEMP CALIBRATION: >AMBIENT TEMP 4.2 oC FL
Operation Service Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-87 CALIBRATE AMBIENT RH: REL HUMIDITY: 40.0 %
Contents Thermo Fisher Scientific Model 5030i SHARP Instruction Manual xi Filter Tape Counter ...
Operation Service Menu 3-88 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The Calibrate Barometer Pressure Offset screen allows
Operation Service Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-89 ● ● ● ● The Vacuum/Flow Calibration submenu
Operation Service Menu 3-90 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific under the filter tape from the positive ΔP port on the
Operation Service Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-91 ● ● ● ● The Flow Calibration submenu is use
Operation Service Menu 3-92 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The Manual Flow Calibration screen allows the user to
Operation Service Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-93 THERMAL MASS COEFFICIENT: CURRENTLY: 0.000000
Operation Service Menu 3-94 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific MASS COEFFICIENT: >AUTO MANUAL
Operation Service Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-95 Note This value is set in the factory and should not b
Operation Service Menu 3-96 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific DETECTOR AUTO CAL: HIGH VOLT: 1500 V BETA CNT :
Operation Service Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-97 HIGH VOLTAGE: CURRENTLY: 1500 V SET TO:
Contents xii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Min and Max Instant Nephelometer Concentration Limits ... 3-84Se
Operation Service Menu 3-98 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ● In the Main Menu, choose Service > Nephelometer C
Operation Service Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-99 SOURCE LEVEL SET: IRED: 70 mA SRC LEV
Operation Service Menu 3-100 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ANALOG OUTPUT CAL: ZERO CONNECT METER TO OUTPUT!
Operation Service Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-101 ANALOG INPUT CAL: >INPUT CHANNEL 1 INPUT CHANNEL
Operation Service Menu 3-102 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The Display Pixel Test screen is used to test the LCD
Operation Password Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-103 ● ● ● ● ● The Password menu allows the use
Operation Password Menu 3-104 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The Lock Instrument screen is used to lock the instr
Operation Password Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-105 The Remove Password screen is used to erase the curr
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 4-1 Chapter 4 Calibration The Model 5030i is an instrument that can make accurate
Contents Thermo Fisher Scientific Model 5030i SHARP Instruction Manual xiii Restore Factory Defaults ...
Calibration Pre-Calibration 4-2 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific nearest 0.1 °C. This thermometer should be referenc
Calibration Calibration Procedure Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 4-3 All screens that appear in this chapter have
Calibration Calibration Procedure 4-4 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Assuming the instrument cover has been remov
Calibration Calibration Procedure Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 4-5 CAL BARO PRESSURE SPAN: PRESSURE: 760
Calibration Calibration Procedure 4-6 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 4–1. Differential Pressure and Vacuum
Calibration Calibration Procedure Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 4-7 ● In the Main Menu, choose Service > Pres
Calibration Calibration Procedure 4-8 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The Mass Calibration submenu is used to view
Calibration Calibration Procedure Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 4-9 Note To achieve the most accurate mass calib
Calibration Calibration Procedure 4-10 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific 3. Cut/break the filter tape and remove fro
Calibration Calibration Procedure Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 4-11 ● ● ● ● The span foil calibration
Contents xiv Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Fan Filter Inspection and Cleaning ...
Calibration Calibration Procedure 4-12 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific MASS COEF: 6878.7 FOIL VALUE: 13
Calibration Calibration Procedure Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 4-13 The Nephelometer Calibration submenu is use
Calibration Calibration Procedure 4-14 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ● In the Main Menu, choose Service > Nep
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 5-1 Chapter 5 Preventive Maintenance This chapter describes the periodic maintena
Preventive Maintenance Replacement Parts 5-2 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific See the “Servicing” chapter for a lis
Preventive Maintenance Cleaning the Inlets Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 5-3 Figure 5–1. U.S. EPA PM10 Inlet Onc
Preventive Maintenance Cleaning the Inlets 5-4 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 5–2. European PM10 Inlet PM1
Preventive Maintenance Fan Filter Inspection and Cleaning Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 5-5 Figure 5–3. PM2.5 Sh
Preventive Maintenance Pump Rebuilding 5-6 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Equipment Damage Some internal compone
Preventive Maintenance Filter Tape Replacement Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 5-7 3. Record the reference volumet
Contents Thermo Fisher Scientific Model 5030i SHARP Instruction Manual xv Flow RTD Replacement ...
Preventive Maintenance Greasing the Cam 5-8 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 5–5. Replacing the Filter Tap
Preventive Maintenance External Pump Exhaust Filter Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 5-9 1. Turn the Heater OFF: Fr
Preventive Maintenance Cleaning the SHARP Optics Assembly 5-10 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Use the following p
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 6-1 Chapter 6 Troubleshooting This instrument has been designed to achieve a high
Troubleshooting Troubleshooting Guides 6-2 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Table 6–1. Troubleshooting - General Gui
Troubleshooting Troubleshooting Guides Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 6-3 Malfunction Possible Cause Action LC
Troubleshooting Troubleshooting Guides 6-4 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Malfunction Possible Cause Action No S
Troubleshooting Troubleshooting Guides Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 6-5 Table 6–2. Troubleshooting - Alarm Mess
Troubleshooting Troubleshooting Guides 6-6 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Alarm Message Possible Cause Action S
Troubleshooting Troubleshooting Guides Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 6-7 Alarm Message Possible Cause Action Pr
© 2011 Thermo Fisher Scientific Inc. All rights reserved. Specifications, terms and pricing are subject to change. Not all products are available in
Contents xvi Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Analog Voltage Outputs ...
Troubleshooting Board-Level Connection Diagrams 6-8 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific 2 PinDI GI T ALOUT P UT S24VDCP
Troubleshooting Board-Level Connection Diagrams Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 6-9 Figure 6–2. Board-Level Conn
Troubleshooting Connector Pin Descriptions 6-10 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 6–3. Board-Level Connectio
Troubleshooting Connector Pin Descriptions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 6-11 Connector Label Reference Designato
Troubleshooting Connector Pin Descriptions 6-12 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Connector Label Reference Designato
Troubleshooting Connector Pin Descriptions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 6-13 Connector Label Reference Designato
Troubleshooting Connector Pin Descriptions 6-14 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Connector Label Reference Designato
Troubleshooting Connector Pin Descriptions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 6-15 Connector Label Reference Designato
Troubleshooting Connector Pin Descriptions 6-16 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Connector Label Reference Designato
Troubleshooting Connector Pin Descriptions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 6-17 Connector Label Reference Designato
Contents Thermo Fisher Scientific Model 5030i SHARP Instruction Manual xvii Communications Configuration ...
Troubleshooting Connector Pin Descriptions 6-18 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Connector Label Reference Designato
Troubleshooting Connector Pin Descriptions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 6-19 Connector Label Reference Designato
Troubleshooting Connector Pin Descriptions 6-20 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Table 6–5. Front Panel Board Connec
Troubleshooting Connector Pin Descriptions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 6-21 Connector Label Reference Designato
Troubleshooting Connector Pin Descriptions 6-22 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Connector Label Reference Designato
Troubleshooting Connector Pin Descriptions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 6-23 Table 6–7. Digital Output Board Co
Troubleshooting Connector Pin Descriptions 6-24 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Connector Label Reference Designato
Troubleshooting Service Locations Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 6-25 Connector Label Reference Designator Pin Si
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-1 Chapter 7 Servicing This chapter explains how to replace the Model 5030i suba
Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
Servicing Service Locations 7-2 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ● “Pressure Board Calibration” on page 7-20 ● “An
Servicing Safety Precautions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-3 Read the safety precautions before beginning any
Servicing Firmware Updates 7-4 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The LCD polarizing plate is very fragile, handle it
Servicing Replacement Parts List Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-5 Table 7-1 lists the replacement parts for the
Servicing Replacement Parts List 7-6 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Part Number Description 10-001403 External P
Servicing Cable List Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-7 Table 7–2 lists the Model 5030i cables. See the “Troubles
Servicing External Device Connection Components 7-8 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Table 7–3 lists the standard a
Servicing External Device Connection Components Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-9 Front Panel Board Pressure Boar
Servicing External Device Connection Components 7-10 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific SHARP Board Bracket AssyRefere
Servicing Removing the SHARP Optics Assembly Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-11 The SHARP optics assembly can be
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual xix Figures Figure 1–1. SHARP Monitor Sample Path ...
Servicing Removing the Measurement Case Assembly and Lowering the Partition Panel 7-12 Model 5030i SHARP Instruction Manual Thermo Fisher Scienti
Servicing Removing the Measurement Case Assembly and Lowering the Partition Panel Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7
Servicing Fuse Replacement 7-14 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Use the following procedure to replace the fuse. E
Servicing Detector Amplifier Assembly Replacement Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-15 5. Remove the four fan moun
Servicing Detector Amplifier Assembly Replacement 7-16 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific 2. Remove the SHARP optics
Servicing Detector Amplifier Board Calibration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-17 Note If Service Mode is not di
Servicing Pressure Board Replacement 7-18 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific 4. Install the new pump by following the
Servicing Pressure Board Replacement Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-19 3. Disconnect the pressure board from th
Servicing Pressure Board Calibration 7-20 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 7–10. Replacing the Pressure Boar
Servicing Analog Output Testing Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-21 b. Turn on manometer and allow to self calibr
Figures xx Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 7–7. Replacing the Detector Amplifier Assembly ...
Servicing Analog Output Testing 7-22 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific 1. Connect a meter to the channel to be teste
Servicing Analog Output Testing Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-23 Figure 7–11. Rear Panel Analog Input and Outp
Servicing Analog Output Calibration 7-24 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Input Channel Pin 4 5 5 6 6 7 7 9 8 10 Gr
Servicing Analog Input Calibration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-25 6. Press to return to the previous scree
Servicing Themistor Replacement 7-26 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Use the following procedure to calibrate the
Servicing I/O Expansion Board (Optional) Replacement Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-27 1. Turn the instrument O
Servicing I/O Expansion Board (Optional) Replacement 7-28 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific 2. Unplug the I/O expans
Servicing Digital Output Board Replacement Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-29 Figure 7–14. Rear Panel Board Conn
Servicing Motherboard Replacement 7-30 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific 5. Pop the digital output board off of the
Servicing Measurement Interface Board Replacement Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-31 Use the following procedure
Figures Thermo Fisher Scientific Model 5030i SHARP Instruction Manual xxi Figure 10–6. Radius Tube Adapter Alignment Check ...
Servicing Photo Interrupt Board Replacement 7-32 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 7–15. Replacing the Measu
Servicing Proportional Valve Replacement Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-33 4. Remove board. 5. Re-install the
Servicing Detector Assembly Replacement 7-34 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific 4. Disconnect plumbing. 5. Remove fo
Servicing Detector Assembly Replacement Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-35 WARNING The detector assembly should
Servicing Flow RTD Replacement 7-36 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 7–18. Replacing the Detector Assembly
Servicing Transformer Replacement Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-37 4. Unplug the RTD connector from the measur
Servicing Radius Tube Replacement 7-38 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific 1. Turn the instrument OFF, unplug the powe
Servicing Ambient RH/Temperature Replacement Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-39 1. Turn the instrument OFF, unpl
Servicing Ambient RH/Temperature Replacement 7-40 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Equipment Damage Some internal
Servicing Front Panel Board Replacement Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-41 Figure 7–22. Replacing the Ambient RH
Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
Servicing LCD Module Replacement 7-42 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 7–23. Replacing the Front Panel Boar
Servicing LCD Module Replacement Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-43 Do not wipe the LCD polarizing plate with a d
Servicing Removing the SHARP Optics Cover 7-44 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The SHARP optics cover can be remov
Servicing External Solenoid Valve Replacement Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-45 Use the following procedure to
Servicing SHARP RH/Temp Sensor Replacement 7-46 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Use the following procedure to rep
Servicing SHARP Interface Board Replacement Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-47 Figure 7–26. Disconnecting the R
Servicing SHARP Interface Board Replacement 7-48 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific 4. Unplug all connectors from the
Servicing Delrin Nut Replacement Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-49 Figure 7–28. Replacing the SHARP Optics Bo
Servicing Delrin Nut Replacement 7-50 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific 2. Turn the Heater OFF: From the Main Menu,
Servicing Internal SHARP Cable Assembly Replacement Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-51 Use the following procedu
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual xxiii Tables Table 1–1. Model 5030i Specifications ...
Servicing Case Cable Replacement 7-52 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 7–30. Replacing the Internal SHARP Ca
Servicing External Zeroing Filter Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-53 4. Replace the case cable assembly by follo
Servicing Service Locations 7-54 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 7–32. Replacing the External Zeroing Filte
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 8-1 Chapter 8 System Description This chapter describes the function and location
System Description Hardware 8-2 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Front Panel Board Pressure BoardFront Panel Tube Mo
System Description Hardware Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 8-3 SHARP Board Bracket Assy Reference Detector BoardSH
System Description Hardware 8-4 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The nephelometer assembly uses an 880 nanometer in
System Description Firmware Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 8-5 This PCB assembly is mounted on the floor plate an
System Description Electronics 8-6 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Signals are gathered from the low-level process
System Description Electronics Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 8-7 External connectors include: ● External Accesso
Tables xxiv Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Table B–8. Reply Termination Formats ...
System Description Electronics 8-8 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Connectors located on the measurement interface
System Description I/O Components Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 8-9 The I/O expansion board connects to the moth
System Description I/O Components 8-10 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The analog outputs may be assigned to any me
System Description I/O Components Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 8-11 Sixteen digital inputs are available, which
System Description I/O Components 8-12 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific 7 or 8 data bits 1 or 2 stop bits No, odd, o
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 9-1 Chapter 9 Optional Equipment The Model 5030i is available with the following
Optional Equipment Terminal Block and Cable Kits 9-2 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific connector board. For associate
Optional Equipment Mounting Options Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 9-3 Table 9–2. Color Codes for 25-Pin and 37-P
Optional Equipment Mounting Options 9-4 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 9–1. Bench Mounting
Optional Equipment Mounting Options Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 9-5 Figure 9–2. EIA Rack Mounting
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 1-1 Chapter 1 Introduction The Model 5030i Synchronized Hybrid Ambient Real-time
Optional Equipment Mounting Options 9-6 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 9–3. Retrofit Rack Mounting
Optional Equipment Mounting Options Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 9-7 Figure 9–4. Rack Mount Option Assembly
Optional Equipment Roof Flange Assembly 9-8 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific An optional roof flange assembly can b
Optional Equipment Roof Flange Assembly Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 9-9 The flange design also has four holes t
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 10-1 Chapter 10 Model 5014i to 5030i Upgrade Procedures This chapter describes th
Model 5014i to 5030i Upgrade Procedures SHARP Connection Bracket Assembly Installation 10-2 Model 5030i SHARP Instruction Manual Thermo Fisher Sc
Model 5014i to 5030i Upgrade Procedures SHARP Connection Bracket Assembly Installation Thermo Fisher Scientific Model 5030i SHARP Instruction Manual
Model 5014i to 5030i Upgrade Procedures SHARP Connection Bracket Assembly Installation 10-4 Model 5030i SHARP Instruction Manual Thermo Fisher Sc
Model 5014i to 5030i Upgrade Procedures Radius Tube Adapter Installation Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 10-5 CONN
Thermo Fisher Scientific WEEE Compliance WEEE Compliance This product is required to comply with the European Union’s Waste Electrical & Electro
Introduction Principle of Operation 1-2 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ● Diaphragm Vacuum Pump ● External Ambien
Model 5014i to 5030i Upgrade Procedures Radius Tube Adapter Installation 10-6 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific 1 =
Model 5014i to 5030i Upgrade Procedures SHARP External Valve Installation Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 10-7 Use
Model 5014i to 5030i Upgrade Procedures SHARP External Valve Installation 10-8 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific c.
Model 5014i to 5030i Upgrade Procedures SHARP Optics Assembly Installation Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 10-9 Us
Model 5014i to 5030i Upgrade Procedures Upgrade of the Heater Tube Assembly 10-10 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
Model 5014i to 5030i Upgrade Procedures Model Label Change Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 10-11 Figure 10–12. Con
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual A-1 Appendix A Warranty Seller warrants that the Products will operate or perform
Warranty Warranty A-2 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific they were not designed, (v) causes external to the Products s
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-1 Appendix B C-Link Protocol Commands This appendix provides a description of t
Introduction Principle of Operation Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 1-3 For details of the analyzer’s principle of
C-Link Protocol Commands Commands B-2 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific For example, if the instrument ID is set to 2
C-Link Protocol Commands Commands Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-3 Table B–1. Error Response Messages Command Re
C-Link Protocol Commands Commands B-4 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Command Description Page alarm amb rh min
C-Link Protocol Commands Commands Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-5 Command Description Page alarm filter tape
C-Link Protocol Commands Commands B-6 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Command Description Page alpha eff Reports
C-Link Protocol Commands Commands Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-7 Command Description Page copy lrec to sp S
C-Link Protocol Commands Commands B-8 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Command Description Page erec layout Repor
C-Link Protocol Commands Commands Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-9 Command Description Page list var log Repo
C-Link Protocol Commands Commands B-10 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Command Description Page no or srec Repor
C-Link Protocol Commands Measurements Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-11 Command Description Page tape counter
Introduction Principle of Operation 1-4 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The SHARP optical assembly senses the light
C-Link Protocol Commands Measurements B-12 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific flags This reports 8 hexadecimal digits
C-Link Protocol Commands Measurements Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-13 Send: neph 24avg Receive: neph 24avg
C-Link Protocol Commands Alarms B-14 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Send: pm 24avg Receive: pm 24avg 10.000E+00
C-Link Protocol Commands Alarms Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-15 alarm amb temp min alarm amb temp max These co
C-Link Protocol Commands Alarms B-16 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific These commands report the current beta count a
C-Link Protocol Commands Alarms Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-17 alarm conc avg pm min alarm conc avg pm max Th
C-Link Protocol Commands Alarms B-18 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Send: alarm conc inst neph min Receive: al
C-Link Protocol Commands Alarms Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-19 representation of the concentration alarm limi
C-Link Protocol Commands Alarms B-20 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Send: set alarm flow pres max 50 Receive:
C-Link Protocol Commands Alarms Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-21 Send: alarm nb status Receive: alarm nb st
Introduction Principle of Operation Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 1-5 HEPA FILTERβ SOURCE Figure 1–2. Nephelomet
C-Link Protocol Commands Alarms B-22 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific alarm neph temp min alarm neph temp max This c
C-Link Protocol Commands Diagnostics Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-23 alarm tape counter max This command repor
C-Link Protocol Commands Diagnostics B-24 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific baro pres This command reports the curren
C-Link Protocol Commands Diagnostics Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-25 Send: diag volt mb Receive: diag volt
C-Link Protocol Commands Diagnostics B-26 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Send: io bd Receive: io bd yes mass T
C-Link Protocol Commands Datalogging Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-27 tape counter This command reports the num
C-Link Protocol Commands Datalogging B-28 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific set copy lrec to sp set copy srec to sp s
C-Link Protocol Commands Datalogging Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-29 The format is defined within the command
C-Link Protocol Commands Datalogging B-30 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific bb = minutes (01 to 59) oo = month (01 to
C-Link Protocol Commands Datalogging Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-31 set lrec format format set srec format fo
Introduction Principle of Operation 1-6 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific A clean filter spot is introduced by an aut
C-Link Protocol Commands Datalogging B-32 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific These commands report the lrecs and srecs
C-Link Protocol Commands Datalogging Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-33 sp field number This command reports the
C-Link Protocol Commands Calibration B-34 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific stream time This command reports if the s
C-Link Protocol Commands Calibration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-35 set alpha th value This command sets the
C-Link Protocol Commands Calibration B-36 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific set baro mass coef value This command set
C-Link Protocol Commands Calibration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-37 set bkg pm value This command is used to
C-Link Protocol Commands Calibration B-38 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Send: set cal flow span 17.2 Receive:
C-Link Protocol Commands Calibration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-39 flow pres span This command reports the
C-Link Protocol Commands Calibration B-40 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific set flow temp offset value This command s
C-Link Protocol Commands Calibration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-41 Send: set thermal mass coef 20.5 Recei
Introduction Principle of Operation Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 1-7 events, a rear-panel mounted zeroing soleno
C-Link Protocol Commands Calibration B-42 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific neph restore bkg This command will restor
C-Link Protocol Commands Calibration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-43 set neph temp offset value This command s
C-Link Protocol Commands Keys/Display B-44 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific set vac pres span value This command set
C-Link Protocol Commands Measurement Configuration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-45 form to save time in transm
C-Link Protocol Commands Measurement Configuration B-46 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Send: conc unit Receive:
C-Link Protocol Commands Measurement Configuration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-47 filter time This command r
C-Link Protocol Commands Measurement Configuration B-48 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific This command sets the heate
C-Link Protocol Commands Measurement Configuration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-49 These commands turn the pre
C-Link Protocol Commands Measurement Configuration B-50 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Selection μg/m3 mg/m3 6 C
C-Link Protocol Commands Hardware Configuration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-51 set temp std value This comman
Introduction Specifications 1-8 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The beta measurement chamber houses an RTD temperat
C-Link Protocol Commands Hardware Configuration B-52 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Level Contrast Level 5 25% 6
C-Link Protocol Commands Hardware Configuration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-53 Send: set default params Re
C-Link Protocol Commands Communications Configuration B-54 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific This command sets the in
C-Link Protocol Commands Communications Configuration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-55 Note This command canno
C-Link Protocol Commands Communications Configuration B-56 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific set allow mode cmd valu
C-Link Protocol Commands Communications Configuration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-57 This command sets the in
C-Link Protocol Commands Communications Configuration B-58 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific set format format This c
C-Link Protocol Commands Communications Configuration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-59 set instrument id value
C-Link Protocol Commands Communications Configuration B-60 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific set power up mode value
C-Link Protocol Commands I/O Configuration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-61 analog iout range channel This com
Introduction Specifications Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 1-9 Connecting and Interface Cables, and Vacuum Pump 3-
C-Link Protocol Commands I/O Configuration B-62 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Send: analog vout range 2 Receiv
C-Link Protocol Commands I/O Configuration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-63 dout channel This command reports
C-Link Protocol Commands I/O Configuration B-64 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific D to A Function Range 8 Current
C-Link Protocol Commands I/O Configuration Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-65 17 frol 19 beta 28 braw 29
C-Link Protocol Commands Record Layout Definition B-66 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Send: set relay open 1 Re
C-Link Protocol Commands Record Layout Definition Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-67 E - parse a 24-bit floatin
C-Link Protocol Commands Record Layout Definition B-68 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Typically, bitfield extracti
C-Link Protocol Commands Record Layout Definition Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-69 I—Indicates a button which p
C-Link Protocol Commands Record Layout Definition B-70 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ' NO:21f{Code_0 Code
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual C-1 Appendix C MODBUS Protocol This appendix provides a description of the MODBUS
Introduction Specifications 1-10 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Filter Media Sample Filter: Glass fiber filter t
MODBUS Protocol TCP Communication Parameters C-2 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific iSeries instruments support the M
MODBUS Protocol Function Codes Thermo Fisher Scientific Model 5030i SHARP Instruction Manual C-3 The function code is a single byte in length. T
MODBUS Protocol Function Codes C-4 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific outputs follow toward the high end of this byte,
MODBUS Protocol Function Codes Thermo Fisher Scientific Model 5030i SHARP Instruction Manual C-5 Here is an example of a request and response to
MODBUS Protocol Function Codes C-6 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Request Function Code 1 Byte 0x03 or 0x04 S
MODBUS Protocol Function Codes Thermo Fisher Scientific Model 5030i SHARP Instruction Manual C-7 Register Value Hi (12) 0x00 Register Value Lo (
MODBUS Protocol MODBUS Parameters Supported C-8 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Here is an example of a request to
MODBUS Protocol MODBUS Parameters Supported Thermo Fisher Scientific Model 5030i SHARP Instruction Manual C-9 Coil Number Status 7 BETA DETECTO
MODBUS Protocol MODBUS Parameters Supported C-10 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Register Number Variable 23 &
MODBUS Protocol MODBUS Parameters Supported Thermo Fisher Scientific Model 5030i SHARP Instruction Manual C-11 Register Number Variable 91 &
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 2-1 Chapter 2 Installation The following installation procedures for the Model 50
MODBUS Protocol MODBUS Parameters Supported C-12 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Coil Number Action Triggered 106
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual D-1 Appendix D Gesytec (Bayern-Hessen) Protocol This appendix provides a descript
Gesytec (Bayern-Hessen) Protocol TCP Communication Parameters D-2 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific iSeries instrume
Gesytec (Bayern-Hessen) Protocol Block Checksum Characters <BCC> Thermo Fisher Scientific Model 5030i SHARP Instruction Manual D-3 The Blo
Gesytec (Bayern-Hessen) Protocol Gesytec Commands D-4 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific If the received command does
Gesytec (Bayern-Hessen) Protocol Gesytec Commands Thermo Fisher Scientific Model 5030i SHARP Instruction Manual D-5 Sample Data Reply String in r
Gesytec (Bayern-Hessen) Protocol Gesytec Commands D-6 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The Gesytec serial number def
Gesytec (Bayern-Hessen) Protocol Gesytec Commands Thermo Fisher Scientific Model 5030i SHARP Instruction Manual D-7 The following 14 measurement
Gesytec (Bayern-Hessen) Protocol Gesytec Commands D-8 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific See Table D–1 for operating
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual E-1 Appendix E ESM Protocol Commands This appendix provides a description of the
Installation Lifting 2-2 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The inner box that the instrument resides in is an ISTA 1A
ESM Protocol Commands ESM Commands Supported E-2 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Command Description UA beta coun
ESM Protocol Commands ESM Commands Supported Thermo Fisher Scientific Model 5030i SHARP Instruction Manual E-3 Command Description E pump on F f
Installation Acceptance Testing and Startup Procedures Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 2-3 Remove Packing(2 pieces)
Installation Acceptance Testing and Startup Procedures 2-4 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific program. Furthermore, it
Installation Acceptance Testing and Startup Procedures Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 2-5 Note If the Model 5030i
Installation Acceptance Testing and Startup Procedures 2-6 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ● Press to return to
Installation Acceptance Testing and Startup Procedures Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 2-7 Figure 2–2. Model 5030i
Installation Acceptance Testing and Startup Procedures 2-8 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific If the RH sensors are sl
Installation Acceptance Testing and Startup Procedures Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 2-9 Attach the small sample
Installation Setup 2-10 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific nephelometer zero should be conducted. After re-zeroing the
Installation Setup Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 2-11 Figure 2–4. Model 5030i Top View—inches [millimeters] Once
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual i About This Manual This manual provides information about installing, operating,
Installation Setup 2-12 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 2–5. Model 5030i Vertical View Ambient Inlet (PM1
Installation Setup Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 2-13 There is one standard length of heated sample tube at one
Installation Setup 2-14 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 2–6. Heater Assembly 4. Attach a second tubing con
Installation Setup Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 2-15 ambient T/RH sensor assembly. Tighten this adapter against
Installation Setup 2-16 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 2–7. Rear Panel Plumbing Schematic Heater Assemb
Installation Connecting External Devices Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 2-17 Several components are available for
Installation Connecting External Devices 2-18 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 2–8. I/O Termin
Installation Connecting External Devices Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 2-19 The analog outputs are arranged on th
Installation Connecting External Devices 2-20 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 2–10. D/O Termi
Installation Connecting External Devices Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 2-21 Assembled Connector down and spacer.
About This Manual Safety ii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ● Appendix B “C-Link Protocol Commands” provides a des
Installation Startup 2-22 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Before the power is connected, be sure that the front pa
Installation Data Content Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 2-23 set to 60 minute logging with 30% memory and Srecs a
Installation Important Tips 2-24 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific DO ● Operate the instrument in a climate-controll
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-1 Chapter 3 Operation This chapter describes the front panel display screens, f
Operation Pushbuttons 3-2 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Figure 3–1. Front Panel Display CAUTION If the LCD pan
Operation Pushbuttons Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-3 Table 3–1 lists the front panel pushbuttons and their fun
Operation Firmware Overview 3-4 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ANGE abcdefghijklmn BKSP opq
Operation Firmware Overview Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-5 submenu contains related instrument settings. This
Operation Firmware Overview 3-6 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The Power-Up screen is displayed when power is a
Operation Firmware Overview Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-7 The Run screen normally displays the current SH
About This Manual Safety and Equipment Damage Alerts Thermo Fisher Scientific Model 5030i SHARP Instruction Manual iii Alert Description If
Operation Firmware Overview 3-8 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The Ambient Conditions Run screen displays the
Operation Firmware Overview Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-9 The Hybrid Data Run screen displays the mass con
Operation Range Menu 3-10 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific XXLO WCONCENTRATIONXXLO W SHARP 15.0 3µg
Operation Range Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-11 CONC UNITS: CURRENTLY: mg/m3 SET TO:
Operation Range Menu 3-12 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The analog outputs are arranged on the rear panel connect
Operation Integration Time Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-13 ● In the Main Menu, choose Range > Set Custom R
Operation 24-Hour Average 3-14 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The 24-Hour Average screen defines a start time dur
Operation Calibration Factors Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-15 CALIBRATION FACTORS: >PM BKG
Operation Calibration Factors Menu 3-16 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The SHARP Background screen allows the use
Operation Calibration Factors Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-17 NEPHELOMETER BKG: >NEPH MODE VALUES R
About This Manual Electrical/Safety Certifications iv Model 5030i SHARP Instruction Manual Thermo Fisher Scientific Alert Description use a so
Operation Instrument Controls Menu 3-18 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific RESTORE PREV BKG VALUES?
Operation Instrument Controls Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-19 INSTRUMENT CONTROLS: >SET FLOW/PUMP SET
Operation Instrument Controls Menu 3-20 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific PUMP: CURRENTLY: ON SET TO:
Operation Instrument Controls Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-21 RH THRESHOLD: CURRENTLY: 65 %
Operation Instrument Controls Menu 3-22 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific START DATE AND TIME: 01 NOV 2008 00:00 ?
Operation Instrument Controls Menu Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-23 ● In the Main Menu, choose Instrument Cont
Operation Volumetric Conditions 3-24 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The Volumetric Conditions menu allows the use
Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-25 The Pressure Compensation screen is used to t
Operation Volumetric Conditions 3-26 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific containing 32 items once each minute for a wee
Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-27 SELECT START POINT BY: SET TO: # OF RE
About This Manual WEEE Symbol Thermo Fisher Scientific Model 5030i SHARP Instruction Manual v The following symbol and description identify the
Operation Volumetric Conditions 3-28 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ERASE LREC LOG FILE DATA?
Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-29 LREC FIELDS: >FIELD 1 SHARP FIEL
Operation Volumetric Conditions 3-30 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific The Other Measurements screen allows the user
Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-31 Figure 3–5. Common Flags Figure 3–6. Detect
Operation Volumetric Conditions 3-32 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific by “<--” after it. Note that at this point,
Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-33 ● ● ● ● The Configure Datalogging m
Operation Volumetric Conditions 3-34 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ● In the Main Menu, choose Instrument Control
Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-35 SET LREC FLAG STATUS DATA: CURRENTLY: OFF
Operation Volumetric Conditions 3-36 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific BAUD RATE: CURRENTLY: 9600
Operation Volumetric Conditions Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-37 STOP BITS: CURRENTLY: 1