Servicing
Pressure Board Replacement
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-19
3. Disconnect the pressure board from the floor plate on the measurement
interface board.
4. Remove the four pressure board assembly retaining screws and remove
the pressure board assembly (Figure 7–9).
Figure 7–9. Replacing the Pressure Board Assembly
5. To install the pressure board assembly, follow the previous steps in
reverse.
Unhook Pressure Cable from
Interface Board “Press” Board
Remove Plumbing
Remove Screws (4)