Sharp DM-4500 Instrucciones de operaciones Pagina 32

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 164
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 31
32
6. Assembly
Transmitted light and incident light analyzer
• Remove the plug cap on the left side of the
stand.
• Insert the analyzer into the receptacle until it
latches in place (Fig. 30).
Motorized analyzer
• Insert the analyzer cube into the correspond-
ing position on the filter turret as described in
section "6.7 Equipping the incident light turret
disk" → p. 30. For details on the proper
position for the analyzer cube, please refer to
the identification sheet provided.
Fig. 30 Inserting the analyzer
1 The plug cap is replaced with the analyzer.
6.9 DIC prisms
• Insert the objective prism slide into the tube
slot (Fig. 31.1). The code letter must match the
code letter on the objective.
• In the Leica DM5000 B microscope, the DIC
prisms are already installed in the DIC disk
above the objective turret (Fig. 68).
Fig. 31 Inserting the objective prism slide
1 Objective prism slide
1
1
Vista de pagina 31
1 2 ... 27 28 29 30 31 32 33 34 35 36 37 ... 163 164

Comentarios a estos manuales

Sin comentarios