Sharp PC-9300 Guía de usuario

Busca en linea o descarga Guía de usuario para Cuadernos Sharp PC-9300. Sharp PC-9300 User guide Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir

Indice de contenidos

Pagina 1 - User Guide

CS 9300User Guide

Pagina 2 - Rochester NY 14 608

2–4 CS 9300 GENERAL OVERVIEWDigital Sensor LocationFigure 2-3 illustrates the location of the digital sensor of the CS 9300.Figure 2–3 Digital Senso

Pagina 3 - Contents

CS 9300 System User Guide (SM748)_Ed 01 2–5Laser LocationsFigure 2-4 illustrates the location of the lasers of the CS 9300.Figure 2–4 CS 9300 Unit Las

Pagina 4

Control Panel2–6 CS 9300 GENERAL OVERVIEWControl PanelThe control panel is an alphanumeric, digital soft touch console. It allows the operator to co

Pagina 5 - About This Guide

X-Ray Remote Control OverviewCS 9300 System User Guide (SM748)_Ed 01 2–7X-Ray Remote Control OverviewThe x- ray remote control enables you to launch a

Pagina 6 - 1–2 About This Guide

Positioning Accessories and Replacement Parts2–8 CS 9300 GENERAL OVERVIEWPositioning Accessories and Replacement PartsThe following accessories are

Pagina 7 - CS 9300 GENERAL OVERVIEW

Positioning Accessories and Replacement PartsCS 9300 System User Guide (SM748)_Ed 01 2–93D Positioning AccessoriesTable 2-2 lists the 3D positioning a

Pagina 8 - Mobile Components

Positioning Accessories and Replacement Parts2–10 CS 9300 GENERAL OVERVIEWPatient Positioning IndicatorsTable 2-3 lists the patient positioning indi

Pagina 9 - General Functional Components

CS 9300 System User Guide (SM748)_Ed 01 3–1Chapter 3IMAGING SOFTWARE OVERVIEWComputer System RequirementsFor the minimum computer system requirements

Pagina 10 - Digital Sensor Location

Acquisition Interface3–2 IMAGING SOFTWARE OVERVIEWAcquisition Interface Panoramic Acquisition OverviewThe Panoramic Acquisition interface is the mai

Pagina 11 - Laser Locations

Acquisition InterfaceCS 9300 System User Guide (SM748)_Ed 01 3–35 Generator Cooling indicator: Indicates the automatic cooling time (mm:ss) required f

Pagina 12 - Control Panel

NoticeCongratulations on your purchase of the CS 9300. Thank you for your confidence in our products and we will do all in our power to ensure your co

Pagina 13 - X-Ray Remote Control Overview

Acquisition Interface3–4 IMAGING SOFTWARE OVERVIEWProgram PaneThe Program pane enables you to choose different radiological exams. The diagram in th

Pagina 14 -  TMJ x2

Acquisition InterfaceCS 9300 System User Guide (SM748)_Ed 01 3–5Patient PaneThe Patient Pane enables you to choose different patient parameters. The s

Pagina 15 - 3D Positioning Accessories

Acquisition Interface3–6 IMAGING SOFTWARE OVERVIEWParameter PaneThe Parameter pane enables you to choose exposure parameters for the radiological im

Pagina 16 -  For panoramic positioning

Acquisition InterfaceCS 9300 System User Guide (SM748)_Ed 01 3–7“Premium” 3D Acquisition Overview The “Premium” 3D Acquisition interface is the optimu

Pagina 17 - IMAGING SOFTWARE OVERVIEW

Acquisition Interface3–8 IMAGING SOFTWARE OVERVIEW6 Stop button: Stops the unit rotative arm movement.7 Pre-shoot button: Performs a single scout im

Pagina 18 - Acquisition Interface

Acquisition InterfaceCS 9300 System User Guide (SM748)_Ed 01 3–93D Program PaneThe 3D Program pane enables you to choose different radiological exams.

Pagina 19

Acquisition Interface3–10 IMAGING SOFTWARE OVERVIEWFigure 3–8 Ear Program PaneEar exam options:Right ear: Acquisition of right ear.Left ear: Acquisi

Pagina 20 - Program Pane

Acquisition InterfaceCS 9300 System User Guide (SM748)_Ed 01 3–11Figure 3–9 Teeth Program PaneFigure 3–10 Jaw Program PaneTeeth setting options:Teeth

Pagina 21 - Patient Pane

Acquisition Interface3–12 IMAGING SOFTWARE OVERVIEWFigure 3–11 TMJ Program PaneTMJ exam options:Right TMJ: Acquisition of right TMJ.Left TMJ: Acquis

Pagina 22 - Parameter Pane

Acquisition InterfaceCS 9300 System User Guide (SM748)_Ed 01 3–13Patient PaneThe Patient pane enables you to choose different patient parameters. The

Pagina 23

CS 9300 System User Guide (SM748)_Ed 01 iiiContents1—About This GuideConventions in this Guide. . . . . . . . . . . . . . . . . . . . . . . . . . . .

Pagina 24

Acquisition Interface3–14 IMAGING SOFTWARE OVERVIEWParameter PaneThe Parameter pane enables you to choose exposure parameters for the radiological i

Pagina 25 - 3D Program Pane

Acquisition InterfaceCS 9300 System User Guide (SM748)_Ed 01 3–15“Select” 3D Acquisition Overview The “Select” 3D Acquisition interface is the 3D inte

Pagina 26

Acquisition Interface3–16 IMAGING SOFTWARE OVERVIEW6 Stop button: Stops the unit rotative arm movement.7 Pre-shoot button: Performs a single scout i

Pagina 27

Acquisition InterfaceCS 9300 System User Guide (SM748)_Ed 01 3–173D Program PaneThe 3D Program pane enables you to choose different radiological exams

Pagina 28

Acquisition Interface3–18 IMAGING SOFTWARE OVERVIEWFigure 3–16 Jaw Program PaneFigure 3–17 TMJ Program PaneBoth jaws restricted acquisition: Ideal f

Pagina 29

Acquisition InterfaceCS 9300 System User Guide (SM748)_Ed 01 3–19Patient PaneThe Patient pane enables you to choose different patient parameters. The

Pagina 30

Acquisition Interface3–20 IMAGING SOFTWARE OVERVIEWParameter PaneThe Parameter pane enables you to choose exposure parameters for the radiological i

Pagina 31

CS 9300 Imaging System User Guide (SM748)_Ed 01 4–1Chapter 4GETTING STARTEDSwitching on the UnitBefore switching on the unit, check that: The install

Pagina 32

Starting the Imaging Software4–2 GETTING STARTED3 Leave the x-ray room and close the door. For each parameter setting, from the x-ray remote control

Pagina 33

Starting the Imaging SoftwareCS 9300 Imaging System User Guide (SM748)_Ed 01 4–32 Enter the required patient information. The Last name, the First nam

Pagina 34

iv ContentsAcquiring a Panoramic, TMJ x2 or Sinus Image . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Pagina 35

Starting the Imaging Software4–4 GETTING STARTEDStarting the CS Imaging SoftwareTo start the CS Imaging Software, follow these steps:1 On your deskt

Pagina 36

Starting the Imaging SoftwareCS 9300 Imaging System User Guide (SM748)_Ed 01 4–54 You can now access the acquisition interface.Accessing the Acquisiti

Pagina 37 - GETTING STARTED

Starting the Imaging Software4–6 GETTING STARTED

Pagina 38 - Starting the Imaging Software

CS 9300 System User Guide (SM748)_Ed 01 5–1Chapter 5ACQUIRING PANORAMIC IMAGESAcquiring a Panoramic, TMJ x2 or Sinus ImageBefore acquiring an image, c

Pagina 39

Acquiring a Panoramic, TMJ x2 or Sinus Image5–2 ACQUIRING PANORAMIC IMAGES5. Press and hold to raise the chin rest to the following positioning in

Pagina 40 - Creating a Patient Card

Acquiring a Panoramic, TMJ x2 or Sinus ImageCS 9300 System User Guide (SM748)_Ed 01 5–3Preparing and Positioning the PatientTo prepare and position th

Pagina 41

Acquiring a Panoramic, TMJ x2 or Sinus Image5–4 ACQUIRING PANORAMIC IMAGES6. On the Control Panel, click to turn ON the 2 positioning laser beams:

Pagina 42 - 4–6 GETTING STARTED

Acquiring a Panoramic, TMJ x2 or Sinus ImageCS 9300 System User Guide (SM748)_Ed 01 5–58. Immobilize the patient head with the temple supports (3). Ti

Pagina 43 - ACQUIRING PANORAMIC IMAGES

Acquiring a Panoramic, TMJ x2 or Sinus Image5–6 ACQUIRING PANORAMIC IMAGESLaunching the X-rayTo launch the x-ray, follow these steps:1. Leave the x-

Pagina 44 -  For a sinus positioning

Acquiring a TMJ x4 ImageCS 9300 System User Guide (SM748)_Ed 01 5–7Acquiring a TMJ x4 ImageBefore acquiring a TMJ x4 image, check that you have: Rese

Pagina 45

CS 9300 System User Guide (SM748)_Ed 01 1–1Chapter 1About This GuideConventions in this GuideThe following special messages emphasize information or i

Pagina 46

X-Ray Dose Emission Information5–8 ACQUIRING PANORAMIC IMAGESX-Ray Dose Emission InformationCompliance with EURATOM 97/43 DirectiveYou can right-cli

Pagina 47

CS 9300 System User Guide (SM748)_Ed 01 6–1Chapter 6ACQUIRING 3D IMAGESAcquiring 3D ImagesBefore acquiring 3D images, check that you have: Reset the

Pagina 48 - Launching the X-ray

Acquiring 3D Images6–2 ACQUIRING 3D IMAGES4. Position the 3D head rest or 3D bite block making sure that it clicks fully into place. 5. Press and ho

Pagina 49 - Acquiring a TMJ x4 Image

Acquiring 3D ImagesCS 9300 System User Guide (SM748)_Ed 01 6–3Preparing and Positioning the PatientTo prepare and position the patient, follow these s

Pagina 50

Acquiring 3D Images6–4 ACQUIRING 3D IMAGES5. Position the patient: There are two possible patient positioning options for 3D examinations:3D Bite Bl

Pagina 51 - ACQUIRING 3D IMAGES

Acquiring 3D ImagesCS 9300 System User Guide (SM748)_Ed 01 6–58. Press and hold to align the upper or lower jaw with the 3D FoV (2) positioning lase

Pagina 52 - 6–2 ACQUIRING 3D IMAGES

Acquiring 3D Images6–6 ACQUIRING 3D IMAGESLaunching the x-rayTo launch the x-ray, follow these steps:1. Leave the x-ray room and close the door. You

Pagina 53

X-Ray Dose Emission InformationCS 9300 System User Guide (SM748)_Ed 01 6–7X-Ray Dose Emission InformationCompliance with EURATOM 97/43 DirectiveYou ca

Pagina 54 - 3D Bite Block method

X-Ray Dose Emission Information6–8 ACQUIRING 3D IMAGES

Pagina 55 - 3D Head Rest method

CS 9300 System User Guide (SM748)_Ed 01 7–1Chapter 7MAINTENANCECleaning ScheduleThis section describes the cleaning schedule that you need to perform

Pagina 56 - Launching the x-ray

Conventions in this Guide1–2 About This Guide

Pagina 57

Monthly7–2 MAINTENANCEMonthlyWipe the outer covers of the unit with a soft and dry cloth.AnnuallyWe recommend a general inspection of the unit carri

Pagina 58 - 6–8 ACQUIRING 3D IMAGES

Controlling the Image QualityDRAFT CS 9300 System User Guide (SM748)_Ed 01 7–33. In the Too l s pane, double-click Quality Assurance Control. The Imag

Pagina 59 - MAINTENANCE

Controlling the Image Quality7–4 MAINTENANCE

Pagina 60 - Controlling the Image Quality

CS 9300 System User Guide (SM748)_Ed 01 8–1Chapter 8TROUBLESHOOTINGQuick TroubleshootingOccasionally, malfunctions can occur during use in the event o

Pagina 61

Quick Troubleshooting8–2 TROUBLESHOOTINGI 6 Wrong rotative arm positionThe exposure button of the x-ray remote control is inactive because the rotat

Pagina 62 - 7–4 MAINTENANCE

CS 9300 System User Guide (SM748)_Ed 01 2–1Chapter 2CS 9300 GENERAL OVERVIEWThe CS 9300 is composed of the following functional components: The unit

Pagina 63 - TROUBLESHOOTING

2–2 CS 9300 GENERAL OVERVIEWMobile ComponentsFigure 2-1 illustrates the up and down movement of the CS 9300 mobile components and the 180° rotation

Pagina 64 - 8–2 TROUBLESHOOTING

CS 9300 System User Guide (SM748)_Ed 01 2–3General Functional ComponentsFigure 2-2 illustrates the general functional components of the CS 9300.Figure