EXT Split Flow Compound Turbomolecular Pumps
INSTALLATION
PAGE
13
Mar 07 13 Issue F
3.3 Connection to the vacuum system
Carefully remove the pump from packaging media and connect to vacuum system.
3.3.1 Mechanical fixing
For variants listed under the Declaration of Incorporation refer to your own guide lines for mechanical
fixing.
WARNING
Once the pump is removed from the bottom packaging, there is a danger of toppling.
WARNING
Install the pump in the vacuum system before you connect the controller to the power
supply. This will ensure that the pump cannot operate and injure people during
installation.
WARNING
The impeller blades on a Split Flow Pump are very sharp. Care should be taken when
handling the pump not to dislodge or damage the protective inlet screens.
WARNING
Under no circumstances must a part of the human body be exposed to a vacuum.
Qty Description
1EXT Split Flow Pump
2 Inlet-screen (fitted)
2 Inlet seal (trapped o-rings)
1 Inlet-strainer (Hi variants only, fitted in the interstage-port)
1 EXDC160 Controller (fitted)*
1 NW10-1/8 inch BSP male extended vent port adaptor*
1 NW10 Centering ring*
1 NW10 Clamping ring*
1 NW25 Centering ring*
1 NW25 Clamping ring*
* Only supplied with system variants.
Table 5 - List of items supplied
WARNING
Ensure all Split Flow Pumps are securely fixed to the vacuum system via their inlet flange
using all available mounting points. If a pump were to seize when not securely mounted,
the stored energy of the rotor could cause rapid movement of the pump, which may
cause damage to equipment and/or injury to personnel.