CarboVis 70x IQ TS Commissioning
3 - 9
ba75950e02 11/2012
3.4.3 Settings for the main sensor
Measuring cycle A measuring cycle consists of the cleaning procedure, the adjustment
time for the measuring system and the determination of the measured
value. The following graphic demonstrates the relevant settings:
Fig. 3-6 Measuring cycle of the UV-VIS sensor
Carrying out settings Using <S>, switch from the measured value display to the main menu
of the settings. Then navigate to the setting menu (setting table) of the
sensor. The procedure is described in detail in your IQ S
ENSOR NET
system operating manual.
A change of the following settings will take some time (up to a few min-
utes) until it can be seen in the measured value display:
Measuring mode
Measuring location
Number of C-sensors
Further changes can only be made when the last changes are visible
in the measured value display.
Default values are marked in bold.