CarboVis 70x IQ TS Commissioning
3 - 7
ba75950e02 11/2012
3.4 Initial commissioning
3.4.1 General information
CAUTION
Never look into the measuring window or put objects into the measur-
ing window during operation! The emitted UV radiation can damage the
eyes!
In the case of authorized use, inadvertent eye contact with the light
beam is not possible.
Measuring location The sensor is prepared for operation in the outlet of the waste water
treatment plant and calibrated in the factory. For use in the inlet you
only have to change the
Measuring location
setting.
The sensor is immediately ready to measure.
How to adjust the settings is described in the following section (section
3.4.3).
User calibration after initial
commissioning
On the basis of a reference measurement ("laboratory value"), check
whether the default settings deliver sufficiently accurate measured
values or whether you want to carry out a user calibration (see section
4.3 C
ALIBRATION).